ABSTRACT
BACKGROUND: Rapid testing for COVID-19 has been clearly identified as an essential component of the strategy to control the SARS-CoV-2 epidemic, worldwide. The ID NOW COVID-19 assay is a simple, user-friendly, rapid molecular biology test based on nicking and extension amplification reaction (NEAR). OBJECTIVES: The aim of this study was to evaluate the ID NOW COVID-19 assay when used as a point-of-care test (POCT) in our Emergency Department (ED). TYPE OF STUDY: This prospective study enrolled 395 consecutive patients; paired nasopharyngeal swabs were collected from each study participant. The first swab was tested with the ID NOW COVID-19 assay at the point-of-care by ED nurses. The second swab was diluted in viral transport medium (VTM) and sent to the clinical microbiology department for analysis by both the RT-PCR Simplexa test COVID-19 Direct assay as the study reference method, and the ID NOW COVID-19 assay performed in the laboratory. RESULTS: Nasopharyngeal swabs directly tested with the ID NOW COVID-19 assay yielded a sensitivity, specificity, PPV and NPV of 98.0%, 97.5%, 96.2% and 98.7%, respectively, in comparison with the RT-PCR study reference assay. When the ID NOW COVID-19 assay was performed in the laboratory using the VTM samples, the sensitivity decreased to 62.5% and the NPV to 79.7%. Three false negative test results were reported with the ID NOW COVID-19 assay when performed using undiluted swabs directly in the ED; these results were obtained from patients with elevated CT values (> 30). CONCLUSION: We demonstrated that the ID NOW COVID-19 assay, performed as a point of care test in the ED using dry swabs, provides a rapid and reliable alternative to laboratory-based RT-PCR methods.
Subject(s)
COVID-19 , COVID-19 Testing , Emergency Service, Hospital , Humans , Nasopharynx , Point-of-Care Testing , Prospective Studies , SARS-CoV-2 , Sensitivity and SpecificityABSTRACT
STUDY OBJECTIVE: Rapid point-of-care (POC) SARS-CoV-2 detection with Abbott ID NOW™ COVID-19 test has been implemented in our Emergency Department (ED) for several months. We aimed to evaluate the operational impact and potential benefits of this innovative clinical pathway. METHODS: We conducted a prospective, descriptive, interventional, non-randomized study, before-after trial with the comparison of patient cohorts from two consecutive periods of seven weeks (observational pre-POC period vs interventional POC period). RESULTS: In 2020, throughout weeks 37 to 50, 3333 patients were assessed for eligibility and among them 331 (9.9%) were positive for SARS-CoV-2 infections. Among the included patients, 136 (9.2%) were positive for SARS-CoV-2 infection in the pre-POC period and 195 (10.5%) in the POC period. Among positive patients for SARS-CoV-2 related infection in-hospital mortality rate was similar between the two groups but the hospitalization rate was higher in the POC group (81.6% vs. 65.4%; p < 0.001). More patients in the POC period were able to leave the ED within 6 h. We examined rates of antibiotic, anticoagulant, and corticosteroid prescriptions among patients tested for SARS-CoV-2 in the ED. Only the rate of prescribed anticoagulants was found to be higher in the POC period (40% vs. 24.2%; p < 0.003). CONCLUSION: We demonstrated that COVID-19 point-of-care testing speeds up clinical decision-making, improving use of recommended treatments for COVID-19, such as anticoagulants. Moreover, it improves the boarding time and significantly shortened the length of stay in the ED for patients requiring outpatient care.